MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2460/ajvr.78.4.508

Медицинская библиотека · American journal of veterinary research · 2017

Effects of repeated use and resterilization on structural and functional integrity of microwave ablation antennas.

Finck Cyrielle A, Zur Linden Alex R, Singh Ameet, Foster Robert A, Nykamp Stephanie G, Sears William C

СтатьяПеревод в очереди

Аннотация

OBJECTIVE To determine effects of repeated use and resterilization on structural and functional integrity of microwave ablation (MWA) antennas. SAMPLE 17 cooled-shaft MWA antennas (3 groups of 5 antennas/group and 2 control antennas). PROCEDURES 1, 2, and 3 ablations in the livers of bovine cadavers were performed at the maximum recommended settings. Antennas were cleaned and sterilized in hydrogen peroxide plasma, and the process was repeated (reprocessing cycle; n = 6). Control antennas were only sterilized (6 times). Aerobic and anaerobic bacterial cultures were performed, and antennas were microscopically assessed for damage. RESULTS 6 cycles were completed. Thirteen of 15 MWA antennas remained functional for up to 4 cycles, 10 were functional after 5 cycles, and only 7 were functional after 6 cycles. Progressive tearing of the silicone coating of the antennas was observed, with a negative effect of the number of cycles for silicone tearing. Size of the ablation zone decreased mildly over time after cycles 5 and 6; however, this was not considered clinically relevant. No significant changes in the shape of ablation zones were detected. All cultures yielded negative results, except for an isolated case, which was considered a contaminant. CONCLUSIONS AND CLINICAL RELEVANCE Structural and functional integrity of the microwave antennas remained acceptable during repeated use and reprocessing for up to 4 cycles. However, there was a decrease in functional integrity at cycles 5 and 6. We suggest that these microwave antennas be subjected to > 3 reprocessing cycles. Antennas should be carefully examined before reuse.

Для цитирования: Finck Cyrielle A, Zur Linden Alex R, Singh Ameet, Foster Robert A, Nykamp Stephanie G, Sears William C. Effects of repeated use and resterilization on structural and functional integrity of microwave ablation antennas.. American journal of veterinary research. 2017. DOI: 10.2460/ajvr.78.4.508.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.