Discordant relationship between Essure microinsert position and tubal occlusion.
Аннотация
Hysteroscopic sterilisation with Essure requires confirmation of tubal occlusion by hysterosalpingogram or microinsert position by transvaginal sonography 3 months after placement before women can rely on the method for pregnancy prevention. A 39-year-old woman underwent hysteroscopic sterilisation via Essure, with successful bilateral tubal occlusion documented on hysterosalpingogram. She had a subsequent unintended pregnancy and termination, and presented with persistent pelvic pain and other non-specific symptoms. She underwent a laparoscopic-assisted vaginal hysterectomy with bilateral salpingectomy, with complete resolution of her symptoms. Pathological evaluation demonstrated a perforated Essure microinsert and ipsilateral tubal occlusion, and a correctly placed Essure microinsert with ipsilateral tubal patency. Clinicians should be cautious about the assumption that correctly placed microinserts based on ultrasonography, hysterosalpingogram or laparoscopic evaluation assures occlusion success.