DOI 10.5005/jp-journals-10024-1845

Медицинская библиотека · The journal of contemporary dental practice · 2016

Mini-implants for Orthodontic Anchorage: Surface Analysis after Redrilling and Sterilization - An in vitro Study.

Gross J M, Nascimento G G, Araújo V C, Bönecker Mjs, Furuse C

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

This study aimed to investigate, in vitro, possible alterations on mini-implants surface after retrieval and if the cleaning process and sterilization can predispose damages.

MATERIALS AND

METHODS

Two commercial mini-implants were tested for deformations after drilling and removing in artificial bone four times. Samples were analyzed by scanning electron microscopy, and surface alterations verified through thread and pitches deformation. To alterations caused by insertion/removal and the cleaning process and sterilization were verified in different procedures: Insertions and sterilization, only insertions, and only sterilization. Photomicrographs were analyzed in order to compare the surface characteristics. Head deformation was verified qualitatively. For a quantitative analysis, distances between threads were measured across the active part of the mini-implants.

RESULTS

No deformation was observed in both groups. The cleaning and sterilization processes did not provoke alteration in both groups. Nevertheless, the presence of synthetic bone was noted in some samples. The mean distances between implant threads were similar after all steps in all regions in both groups.

CONCLUSION

The results suggest that the tested mini-implants can be retrieved without damage of its surface after four cycles of insertion, removal, and sterilization.

KEYWORDS: Orthodontic mini-implant, Redrilling, Sterilization.

CLINICAL SIGNIFICANCE: Mini-implants can be retrieved without damage to its surface after four cycles of insertion, removal, and sterilization in the same patient without representing a biological concern.

Для цитирования: Gross J M, Nascimento G G, Araújo V C, Bönecker Mjs, Furuse C. Mini-implants for Orthodontic Anchorage: Surface Analysis after Redrilling and Sterilization - An in vitro Study.. The journal of contemporary dental practice. 2016. DOI: 10.5005/jp-journals-10024-1845.