Diving and hyperbaric medicine · 2023
An evaluation of the NUI Compact Chest Compression Device (NCCD), a mechanical CPR device suitable for use in the saturation diving environment.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Tabner Andrew, Bryson Philip, Tilbury Nicholas, McGregor Benjamin, Wesson Alistair, Hughes Gareth D, Hughes Gareth R, Johnson Graham
Аннотация
INTRODUCTION: Provision of manual chest compressions in a diving bell using a conventional technique is often impossible, and alternative techniques are poorly evidenced in terms of efficacy and sustainability. The first mechanical cardiopulmonary resuscitation (CPR) device suitable for use in this environment, the NUI Compact Chest Compression Device (NCCD), has recently been designed and manufactured. This study assessed both the efficacy of the device in delivering chest compressions to both prone and seated manikins, and the ability of novice users to apply and operate it.
METHODS: Compression efficacy was assessed using a Resusi Anne QCPR intelligent manikin, and the primary outcome was the proportion of compressions delivered to target depth (50-60 mm). The gold standard was that achieved by expert CPR providers delivering manual CPR; the LUCAS 3 mCPR device was a further comparator.
RESULTS: The NCCD delivered 100% of compressions to target depth compared to 98% for the gold standard (interquartile range 1.5%) and 98% for the LUCAS 3 when applied to both supine and seated manikins. The NCCD sometimes became dislodged and had to be reapplied when used with a seated manikin.
CONCLUSIONS: The NCCD can deliver chest compressions at target rate and depth to both supine and seated manikins with efficacy equivalent to manual CPR and the LUCAS 3. It can become dislodged when applied to a seated manikin; its design has now been altered to prevent this. New users can be trained in use of the NCCD quickly, but practise is required to ensure effective use.
METHODS: Compression efficacy was assessed using a Resusi Anne QCPR intelligent manikin, and the primary outcome was the proportion of compressions delivered to target depth (50-60 mm). The gold standard was that achieved by expert CPR providers delivering manual CPR; the LUCAS 3 mCPR device was a further comparator.
RESULTS: The NCCD delivered 100% of compressions to target depth compared to 98% for the gold standard (interquartile range 1.5%) and 98% for the LUCAS 3 when applied to both supine and seated manikins. The NCCD sometimes became dislodged and had to be reapplied when used with a seated manikin.
CONCLUSIONS: The NCCD can deliver chest compressions at target rate and depth to both supine and seated manikins with efficacy equivalent to manual CPR and the LUCAS 3. It can become dislodged when applied to a seated manikin; its design has now been altered to prevent this. New users can be trained in use of the NCCD quickly, but practise is required to ensure effective use.