Effectiveness of the SYSTEM 1E Liquid Chemical Sterilant Processing System for reprocessing duodenoscopes.
Аннотация
BACKGROUND
A troubling number of health care-acquired infection outbreaks and transmission events, some involving highly resistant microbial pathogens and resulting in serious patient outcomes, have been traced to reusable, high-level disinfected duodenoscopes in the United States. The Food and Drug Administration (FDA) requested a study be conducted to verify liquid chemical sterilization efficacy of SYSTEM 1E(®) Liquid Chemical Sterilant Processing System (STERIS Corporation, Mentor, OH) with varied duodenoscope designs under especially arduous conditions. Here, we describe the system's performance under worst case SYSTEM 1E(®) processing conditions.
METHODS
The test protocol challenged the system's performance by running a fractional cycle to evaluate reduction of recoverable test spores from heavily contaminated endoscopes, including all channels and each distal tip, under worst case SYSTEM 1E(®) processing conditions.
RESULTS
All devices were successfully liquid chemically sterilized, showing greater than a 6 log10 reduction of Geobacillus stearothermophilus spores at every inoculation site of each duodenoscope tested, in less than half the exposure time of the standard cycle.
CONCLUSIONS
The successful outcome of the additional efficacy testing reported here indicates that the SYSTEM 1E(®) is an effective low-temperature liquid chemical sterilization method for duodenoscopes and other critical and semicritical devices. It offers a fast, safe, convenient processing alternative while providing the assurance of a system expressly tested and cleared to achieve liquid chemical sterilization of specific validated duodenoscope models.