Impact of a High-Power Light-Emitting Diode Unit With Reduced Curing Times on the Shear Bond Strength of Orthodontic Brackets and Their Adhesive Remnant Index Scores.

Rohini Neelapala, Sunil Ghanta, Ram Rsvm Raghu, Ranganayakulu Inuganti, Vijaya Krishna Balanaga

СтатьяПеревод в очереди

Аннотация

AIM AND

OBJECTIVES

This study aims to evaluate the curing time minimally required for bonding stainless-steel (SS) brackets using a high-power light-emitting diode (LED) light curing unit (LCU) and examine the debonded enamel surface for adhesive remnant.

MATERIALS AND

METHODS

Based on the LED LCU and curing time employed, 80 human maxillary first premolar teeth were equally segregated into four groups. Three groups were cured using a high-power LED unit (Guilin Woodpecker Medical Instrument Co., Ltd., Guilin, Guangxi, China) for one, two, and three seconds. The fourth group served as a control and was bonded with another intensive LED unit (Elipar™ S10 LED Curing Light; 3M, Saint Paul, Minnesota, United States) for 20 seconds. Transbond™ XT Light Cure Adhesive (3M, United States) adhesive was used for bonding the SS brackets. All the samples were exposed to shear bond strength (SBS) testing after a 24-hour immersion period in distilled water at 37°C. A stereomicroscope was used to examine and score the adhesive remnant on the debonded surface using a modified adhesive remnant index (ARI). Kruskal-Wallis-ANOVA and post-hoc Mann-Whitney U tests for multiple pairwise comparisons were performed to analyze the data.

RESULTS

Time and intensity significantly affected the SBS (P<0.001). A higher SBS value (16.04 megapascals (MPa)) was obtained in the six-second group when compared to the three-second (11.58 MPa), one-second (10.69 MPa), and 20-second control (13 MPa) groups. The ARI was significantly affected by the curing method.

CONCLUSIONS

Higher SBSs were recorded for the six-second group using the high-power LED. A greater ARI score is associated with a reduced curing duration and vice versa.

Для цитирования: Rohini Neelapala, Sunil Ghanta, Ram Rsvm Raghu, Ranganayakulu Inuganti, Vijaya Krishna Balanaga. Impact of a High-Power Light-Emitting Diode Unit With Reduced Curing Times on the Shear Bond Strength of Orthodontic Brackets and Their Adhesive Remnant Index Scores.. Cureus. 2023. DOI: 10.7759/cureus.39855.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.