MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s0002-8703(77)80250-0

Медицинская библиотека · American heart journal · 1977

Reoperation for dysfunction of the Björk-Shiley mitral disc prosthesis: report of eight cases.

Maroñas J M, Rufilanchas J J, Villagra F, Tellez G, Agosti J, Figuera D

СтатьяПеревод в очереди

Аннотация

Of 193 patients with Björk-Shiley mitral valve prostheses, replacement was necessary in 8 (4.1%). The reasons for reoperation were: detachment (4), thrombosis (1), technical error (1), and late disc entrapment (2). Five of these patients died (62.5%), the death being directly related to the need for urgent operation because of low cardiac output. We recommend avoiding the use of the larger sized Björk-Shiley prostheses, since striking of the disc against the ventricle wall, probably consequent to postoperative decrease in heart size, may appear even 1 year after implantation of the prosthesis. An early diagnosis and early reoperation offer these patients a much more favorable prognosis.

Для цитирования: Maroñas J M, Rufilanchas J J, Villagra F, Tellez G, Agosti J, Figuera D. Reoperation for dysfunction of the Björk-Shiley mitral disc prosthesis: report of eight cases.. American heart journal. 1977. DOI: 10.1016/s0002-8703(77)80250-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.