DOI 10.1016/j.mri.2026.110803

Медицинская библиотека · Magnetic resonance imaging · 2026

RF safety assessment of dental retainers for human head MRI array coils at 9.4 T.

Berezko Egor V, Solomakha Georgiy A, Avdievich Nikolai I, Scheffler Klaus

СтатьяПеревод в очереди

Аннотация

In this work we developed a numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers. The proposed workflow includes evaluation of the B1+ distribution and peak Specific Absorption Rate (pSAR) for the standard circularly polarized modes, worst-case pSAR modes, and random RF modes. The workflow provides a safety margin, which limits the RF transmit power during MRI scanning. The proposed workflow is applicable to different array coils at different UHF strengths (e.g. 7 T, 9.4 T, 11.7 T, 14 T) as well as to other types of conductive objects (e.g., dental or neural implants). 16-element tight-fit transceiver and larger 16-element transmit-only head arrays were numerically evaluated at 9.4 T in the presence of the dental retainer. Updated safety margins were calculated for both arrays. In addition, we calculated Q-matrix and virtual observation point matrices, which facilitate the real-time SAR-monitoring and can be used for optimization of the B1+ homogeneity and SAR.

Для цитирования: Berezko Egor V, Solomakha Georgiy A, Avdievich Nikolai I, Scheffler Klaus. RF safety assessment of dental retainers for human head MRI array coils at 9.4 T.. Magnetic resonance imaging. 2026. DOI: 10.1016/j.mri.2026.110803.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.