Impact of rubber dam thickness on image quality parameters in photostimulable phosphor plates.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Özel Şelale, Amasya Hakan, Yanık Nalbantoğlu Deniz
Аннотация
OBJECTIVE: The aim of this study is to investigate the effects of different rubber dam thicknesses on image quality in 2 exposure protocols.
METHODS: For the study, 3 rubber dam thicknesses were used: thin (0.14 mm), medium (0.18 mm), and heavy (0.22 mm). To mimic clinical conditions, a rubber dam was used in 2 layers. Exposure was performed using 2 different protocols: Protocol 1: 0.080 s exposure time, 65 kV, 7 mA, and Protocol 2: 0.160 s exposure time, 65 kV, 7 mA. For each thickness and protocol, 47 measurements were taken (n = 47). Radiographic images were exported in TIFF and analysed using ImageJ software. The region of interest (ROI) was determined, and signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), Michelson, and Weber contrast values were obtained. One-way ANOVA, post hoc Tukey, and intra-class correlation were used for statistical analysis.
RESULTS: No statistical difference was detected for Protocol 1 between rubber dam thicknesses in SNR, CNR, Michelson, and Weber contrasts (P > .05). For protocol 2 (longer exposure time), heavy and medium thickness had lower SNR and CNR values than the thin one (P < .05). Michelson and Weber contrasts were statistically changed in different thicknesses of rubber dam (P < .05). ICC values were good and excellent.
CONCLUSIONS: The thick rubber dam reduced SNR and CNR values; likewise, Michelson and Weber contrasts were changed, which pointed a reduced image quality and negatively affected object visibility. Short exposure times are recommended to maintain image quality in clinical situations requiring the use of a thick rubber dam.
METHODS: For the study, 3 rubber dam thicknesses were used: thin (0.14 mm), medium (0.18 mm), and heavy (0.22 mm). To mimic clinical conditions, a rubber dam was used in 2 layers. Exposure was performed using 2 different protocols: Protocol 1: 0.080 s exposure time, 65 kV, 7 mA, and Protocol 2: 0.160 s exposure time, 65 kV, 7 mA. For each thickness and protocol, 47 measurements were taken (n = 47). Radiographic images were exported in TIFF and analysed using ImageJ software. The region of interest (ROI) was determined, and signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), Michelson, and Weber contrast values were obtained. One-way ANOVA, post hoc Tukey, and intra-class correlation were used for statistical analysis.
RESULTS: No statistical difference was detected for Protocol 1 between rubber dam thicknesses in SNR, CNR, Michelson, and Weber contrasts (P > .05). For protocol 2 (longer exposure time), heavy and medium thickness had lower SNR and CNR values than the thin one (P < .05). Michelson and Weber contrasts were statistically changed in different thicknesses of rubber dam (P < .05). ICC values were good and excellent.
CONCLUSIONS: The thick rubber dam reduced SNR and CNR values; likewise, Michelson and Weber contrasts were changed, which pointed a reduced image quality and negatively affected object visibility. Short exposure times are recommended to maintain image quality in clinical situations requiring the use of a thick rubber dam.