BACKGROUND

Manufacturers' recommendations on cleaning of tracheostomy tubes focus on general warning information and non-specific manual cleaning procedures. The aim of this experimental study was to evaluate different reprocessing methods and to determine the mechanical integrity and functionality of tracheostomy tubes following reprocessing.

METHODS

Sixteen cuffed or un-cuffed tracheostomy tubes obtained from hospital in-patients were reprocessed using one of the following reprocessing methods: a) manual brushing and rinsing with tap water, b) manual brushing followed by disinfection with a glutaraldehyde solution, c) manual brushing followed machine-based cleaning in a dishwasher, and d) manual brushing followed by ultrasound cleaning in a commercially available ultrasound device. Microbial burden of the tubes before and after reprocessing was assessed by measurement of microbial colony-forming units per mL (CFU/mL) of rinsing fluid. After cleaning, tracheostomy tubes were investigated for loss of functionality.

FINDINGS: Manual brushing and rinsing with tap water reduced microbial colonization in average by 10(2) CFU/mL, but with poor reproducibility and reliability. Complete microbial reduction was achieved only with additional chemical or machine-based thermal disinfection. Ultrasound sonification yielded no further microbial reduction after manual brushing.

CONCLUSION

Manual brushing alone will not result in complete eradication of microorganism colonising cuffed or un-cuffed tracheostomy tubes. However, manual cleaning followed by chemical or thermal disinfection may be regarded as safe and reproducible reprocessing method. If a machine-based reprocessing method is used for cuffed tubes, the cuffs' ventilation hose must be secured in a safe position prior to thermal disinfection.

Для цитирования: Leonhard Matthias, Assadian Ojan, Zumtobel Michaela, Schneider-Stickler Berit. Microbiological evaluation of different reprocessing methods for cuffed and un-cuffed tracheostomy tubes in home-care and hospital setting.. GMS hygiene and infection control. 2016. DOI: 10.3205/dgkh000262.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.