Screening test for detection of metallic foreign objects in the orbit before magnetic resonance imaging.

Otto P M, Otto R A, Virapongse C, Friedman S M, Emerson S, Li K C, Malot R, Kaude J V, Staab E V

СтатьяПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVES

A study was designed to determine whether plain films, used as a screening modality for magnetic resonance imaging (MRI), could reliably detect intraorbital metallic foreign objects.

METHODS

In the first experiment, 20 metal particles were placed in five human cadaver orbits. Routine orbital plain film series and computed tomography (CT) were obtained, randomized, and interpreted blinded by three experienced radiologists.

RESULTS

The threshold size of particle detection for CT (0.07 mm3) was lower than for plain films (0.12 mm3). Placing metal particles in artificial and true vitreous demonstrated that all particles moved under a magnetic field at 1.5 T. When human globes were exposed to industrial tools (grinder, bandsaw, air hose, etc.), no metal objects penetrated the sclera.

CONCLUSIONS

Plain films can be used as a low-cost, low-radiation screening procedure for high-risk patients with occupations involving metal work. CT should be used for patients with a history of eye trauma from other causes.

Для цитирования: Otto P M, Otto R A, Virapongse C, Friedman S M, Emerson S, Li K C, Malot R, Kaude J V, Staab E V. Screening test for detection of metallic foreign objects in the orbit before magnetic resonance imaging.. Investigative radiology. 1992. DOI: 10.1097/00004424-199204000-00010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.