DOI 10.1097/ijg.0000000000002723

Медицинская библиотека · Journal of glaucoma · 2026

Perimetric Outcomes of Melbourne Rapid Field Perimetry in Patients With Glaucoma: A Systematic Review and Meta-Analysis.

Inzunza Andres, Moreno-Diaz Edgar A, Goss Deborah, Friedman David S

СтатьяСистематический обзорМетаанализПеревод в очереди

Аннотация

PRCIS: Melbourne Rapid Fields (MRF) had shorter test durations than HFA, high agreement with HFA global indices, modest bias in MD and PSD measurements, and comparable concordance and repeatability. MRF had poorer reliability indices, while maintaining good diagnostic performance.

PURPOSE: The Humphrey Field Analyzer (HFA) is the most widely used perimeter in glaucoma, but its cost and low portability limit its use in resource-constrained settings. Melbourne Rapid Fields (MRF) is a low-cost, tablet-based perimeter system that enables at-home and low-resource testing. Despite growing evidence that the MRF can obtain reliable estimates of the visual field, no study has synthesized its perimetric performance through meta-analysis and compared it with HFA.

METHODS

Following PROSPERO registration, we conducted a systematic review to compare MRF with HFA in patients with glaucoma. Searches were performed using controlled vocabulary related to perimetry assessment with these devices. Meta-analyses and narrative synthesis were conducted to summarize their performance.

RESULTS

Of 290 studies screened, 9 were included, totaling 777 eyes. Pooled data comparing MRF to HFA, MRF had faster testing (-1.33 min), less-negative MD (+2.08 dB), and higher PSD (+0.77 dB). Global agreement with HFA was high (MD ICC=0.94; MD R2=0.85; PSD ICC=0.85). Test-retest reliability was similar between devices (MRF ICC=0.96; HFA ICC=0.95). AUCs were comparable for MD (0.84 vs. 0.85) but lower for PSD (0.81 vs. 0.93). MRF had higher fixation losses, false positives, and false negatives.

CONCLUSION

Melbourne Rapid Fields (MRF) had shorter test durations, high global agreement with HFA, excellent repeatability, and a bias toward less-negative global results. MRF showed greater pointwise variability at certain thresholds and worse reliability measures. Despite these differences, MRF provided diagnostic information comparable to that of HFA.

Для цитирования: Inzunza Andres, Moreno-Diaz Edgar A, Goss Deborah, Friedman David S. Perimetric Outcomes of Melbourne Rapid Field Perimetry in Patients With Glaucoma: A Systematic Review and Meta-Analysis.. Journal of glaucoma. 2026. DOI: 10.1097/ijg.0000000000002723.