Shetty Rajesh, Chokshi Amit, Paauw Mark, Francis Joanne, Freidl Kathryn, Orey Dontia, Egli Anna, Keeton Haleigh, Click Kaitlyn, Schaffner Hannah, Wagner Isabella, Hasan Akbar, Gayasaddin Omar, Lott McGregor, Goldfaden Rebecca
Аннотация
PURPOSE: Evaluate whether visual acuity is influenced by intraocular lens (IOL) decentration or tilt with Clareon® Vivity® and Clareon® Vivity® Toric IOLs.
METHODS: This was a non-interventional, retrospective, single-arm, observational study. Adults with Clareon® Vivity® or Vivity® Toric IOLs without significant posterior capsule opacification, significant history of ocular surgery, or major ocular pathologies were included. Postoperative assessments performed ≥1 month after surgery included monocular high-contrast corrected and uncorrected distance visual acuity (CDVA, UDVA, 4 m), distance-corrected and uncorrected intermediate visual acuity (DCIVA, UIVA, 66 cm), manifest refraction spherical equivalent (MRSE), low-contrast CDVA, IOL decentration and tilt, mesopic pupil size, and corneal higher-order aberrations. The primary outcome was the association between the CDVA and IOL decentration or tilt.
RESULTS: A total of 202 eyes (n=101) were analyzed: 140 eyes (69%) received Clareon® Vivity® and 62 eyes (31%) received Vivity® Toric. The mean IOL decentration was 0.211 mm (SD 0.13) and the mean tilt was 5.6° (SD = 1.18). Mean CDVA was 0.056 logMAR, UCDVA was 0.155 logMAR, DCIVA was 0.018 logMAR, UCIVA was 0.019 logMAR, and MRSE was -0.147 D. In unadjusted analyses, no significant correlations were found between CDVA and decentration (r = 0.12, p = 0.08) or tilt (r = 0.06, p = 0.43), nor between DCIVA and decentration (r = 0.07, p = 0.34) or tilt (r = -0.13, p = 0.07). In adjusted multivariable regression analyses, IOL decentration was significantly associated with postoperative CDVA100 (β=0.618, p=0.023), whereas residual refractive cylinder and IOL tilt were not independently associated with postoperative visual outcomes.
CONCLUSION: Clareons' Vivity® and Vivity® Toric IOLs demonstrated excellent distance and intermediate visual acuity. Visual outcomes remained generally favorable despite modest decentration or tilt, although adjusted analyses identified a modest association between decentration and postoperative CDVA100, supporting clinical reliability of this IOL design.
METHODS: This was a non-interventional, retrospective, single-arm, observational study. Adults with Clareon® Vivity® or Vivity® Toric IOLs without significant posterior capsule opacification, significant history of ocular surgery, or major ocular pathologies were included. Postoperative assessments performed ≥1 month after surgery included monocular high-contrast corrected and uncorrected distance visual acuity (CDVA, UDVA, 4 m), distance-corrected and uncorrected intermediate visual acuity (DCIVA, UIVA, 66 cm), manifest refraction spherical equivalent (MRSE), low-contrast CDVA, IOL decentration and tilt, mesopic pupil size, and corneal higher-order aberrations. The primary outcome was the association between the CDVA and IOL decentration or tilt.
RESULTS: A total of 202 eyes (n=101) were analyzed: 140 eyes (69%) received Clareon® Vivity® and 62 eyes (31%) received Vivity® Toric. The mean IOL decentration was 0.211 mm (SD 0.13) and the mean tilt was 5.6° (SD = 1.18). Mean CDVA was 0.056 logMAR, UCDVA was 0.155 logMAR, DCIVA was 0.018 logMAR, UCIVA was 0.019 logMAR, and MRSE was -0.147 D. In unadjusted analyses, no significant correlations were found between CDVA and decentration (r = 0.12, p = 0.08) or tilt (r = 0.06, p = 0.43), nor between DCIVA and decentration (r = 0.07, p = 0.34) or tilt (r = -0.13, p = 0.07). In adjusted multivariable regression analyses, IOL decentration was significantly associated with postoperative CDVA100 (β=0.618, p=0.023), whereas residual refractive cylinder and IOL tilt were not independently associated with postoperative visual outcomes.
CONCLUSION: Clareons' Vivity® and Vivity® Toric IOLs demonstrated excellent distance and intermediate visual acuity. Visual outcomes remained generally favorable despite modest decentration or tilt, although adjusted analyses identified a modest association between decentration and postoperative CDVA100, supporting clinical reliability of this IOL design.