MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/mmy/myz064

Медицинская библиотека · Medical mycology · 2020

False negative diagnostic errors with polymerase chain reaction for the detection of cryptococcal meningoencephalitis.

Lewis Paul O, Lanier Cameron G, Patel Paras D, Krolikowski Whitney D, Krolikowski Matthew A

СтатьяПеревод в очереди

Аннотация

The accuracy of the BioFire FilmArray Meningitis/Encephalitis (ME) panel for the identification of Cryptococcus has recently been called into question. The primary objective of this study was to assess the agreement between the BioFire ME polymerase chain reaction (PCR) and other markers of cryptococcal infection. This retrospective review identified five patients with cryptococcal meningoencephalitis, 4 of whom had a negative ME panel for Cryptococcus. All five cases had positive serum cryptococcal antigens, and three of five had a positive cerebrospinal fluid (CSF) culture for Cryptococcus. The BioFire ME panel does not appear to be reliable for ruling out Cryptococcus meningoencephalitis; multiple testing methods are recommended.

Для цитирования: Lewis Paul O, Lanier Cameron G, Patel Paras D, Krolikowski Whitney D, Krolikowski Matthew A. False negative diagnostic errors with polymerase chain reaction for the detection of cryptococcal meningoencephalitis.. Medical mycology. 2020. DOI: 10.1093/mmy/myz064.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.