Modeling and Reproducibility of Twin Concentric Electrical Impedance Myography.
Аннотация
OBJECTIVE
Electrical impedance myography (EIM) is a recent technology to assess muscle health. As of today, the clinical application of EIM has been applied only to evaluate muscle condition using non-invasive surface electrodes in contact with the skin; however, intermediate tissues at the recording site introduce confounding artifacts which reduce the technique's performance as a biomarker of neuromuscular disorders (NMD). Here, we develop and test in humans a new approach using two concentric needles for intramuscular EIM recordings.
METHODS
First, we study the recording characteristics of dual concentric needle EIM via analytical models and finite element models (FEMs). Next, the validity of the models is verified by performing experiments on saline and agar phantoms. Finally, 8 subjects with various neuromuscular diseases were studied measuring tibialis anterior, biceps, deltoid, adductor pollicis brevis, first dorsal interosseous and flexor carpi radialis muscles.
RESULTS
Analytical and FEM simulations are in good agreement with a maximum experimental discrepancy 8% and 9% using gauge needles 26 and 30, respectively. The inter-session reproducibility, as measured by the intraclass correlation coefficients for all muscles studied, was 0.926, which is comparable or exceeds the reproducibility of other well-established electrophysiological tests to assess muscle health.
CONCLUSION
The reproducibility of the technique support future clinical validation of needle EIM for assessment of disease status, either as part of standard patient care or as biomarker measure in clinical trials.
SIGNIFICANCE: Needle EIM has the potential of becoming a valuable diagnostic tool to evaluate NMD in adult population.