Bevel angle study of flexible hollow needle insertion into biological mimetic soft-gel: Simulation and experimental validation.

Jushiddi Mohamed G, Cahalane Rachel M, Byrne Michael, Mani Aladin, Silien Christophe, Tofail Syed A M, Mulvihill John J E, Tiernan Peter

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

A thorough understanding of cutting-edge geometry and cutting forces of hollow biopsy needles are required to optimise needle tip design to improve fine needle aspiration procedures.

OBJECTIVES

To incorporate the dynamics of needle motion in a model for flexible hollow bevel tipped needle insertion into a biological mimetic soft-gel using parameters obtained from experimental work. Additionally, the models will be verified against corresponding needle insertion experiments.

METHODS

To verify simulation results, needle deflection and insertion forces were compared with corresponding experimental results acquired with an in-house developed needle insertion mechanical system. Additionally, contact stress distribution on needles from agar gel for various time scales were also studied.

RESULTS

For the 15°, 30°, 45°, 60° bevel angle needles, and 90° blunt needle, the percentage error in needle deflection of each needle compared to experiments, were 7.3%, 9.9%, 8.6%, 7.8%, and 9.7% respectively. Varying the bevel angle at the needle tip demonstrates that the needle with a lower bevel angle produces the largest deflection, although the insertion force does not vary too much among the tested bevel angles.

CONCLUSION

This experimentally verified computer-based simulation model could be used as an alternative tool for better understanding the needle-tissue interaction to optimise needle tip design towards improved biopsy efficiency.

Для цитирования: Jushiddi Mohamed G, Cahalane Rachel M, Byrne Michael, Mani Aladin, Silien Christophe, Tofail Syed A M, Mulvihill John J E, Tiernan Peter. Bevel angle study of flexible hollow needle insertion into biological mimetic soft-gel: Simulation and experimental validation.. Journal of the mechanical behavior of biomedical materials. 2020. DOI: 10.1016/j.jmbbm.2020.103896.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.