Accuracy of intraocular pressure measurements in dogs using two different tonometers and plano therapeutic soft contact lenses.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ahn Jeong-Taek, Jeong Man-Bok, Park Young-Woo, Kim Se-Eun, Ahn Jae-Sang, Lee Yes-Ran, Lee Eui-Ri, Seo Kangmoon
Аннотация
OBJECTIVE: To compare and evaluate the accuracy of intraocular pressure (IOP) measured through a therapeutic contact lens, using applanation (TonoPen XL(®)) and rebound (TonoVet(®)) tonometers in enucleated dog eyes.
ANIMALS STUDIED: A total of 30 enucleated eyes from 15 beagle dogs.
PROCEDURES: To measure accurate IOP, the anterior chamber of each enucleated eye was cannulated with two 26-gauge needles and two polyethylene tubes were connected vertically to an adjustable reservoir bag of normal saline and a pressure transducer. IOP was measured by the TonoPen XL(®) followed by the TonoVet(®) without a contact lens. After a contact lens was applied to the cornea, IOP was re-measured in the same order. Three consecutive IOP measurements were performed using both tonometers.
RESULTS: Without the contact lens, the IOP values obtained by both tonometers correlated well according to the regression analysis (TonoVet(®): γ(2) = 0.98, TonoPen XL(®): γ(2) = 0.97, P < 0.001). The TonoPen XL(®) consistently underestimated values as transducer IOP increased; however, IOP values measured with the TonoPen XL(®) were in close agreement and were less variable than those determined with the TonoVet(®) when a contact lens was applied to the cornea. Bland-Altman analysis was used to determine the lower and upper limits of agreement (TonoVet(®): -29.7 and +21.1 mmHg, TonoPen XL(®): -3.9 and +3.6 mmHg) between the two devices.
CONCLUSIONS: This study suggests that the TonoPen XL(®) is a useful tonometer for dogs wearing therapeutic contact lenses, and importantly, contact lenses would not need to be removed prior to IOP measurement.
ANIMALS STUDIED: A total of 30 enucleated eyes from 15 beagle dogs.
PROCEDURES: To measure accurate IOP, the anterior chamber of each enucleated eye was cannulated with two 26-gauge needles and two polyethylene tubes were connected vertically to an adjustable reservoir bag of normal saline and a pressure transducer. IOP was measured by the TonoPen XL(®) followed by the TonoVet(®) without a contact lens. After a contact lens was applied to the cornea, IOP was re-measured in the same order. Three consecutive IOP measurements were performed using both tonometers.
RESULTS: Without the contact lens, the IOP values obtained by both tonometers correlated well according to the regression analysis (TonoVet(®): γ(2) = 0.98, TonoPen XL(®): γ(2) = 0.97, P < 0.001). The TonoPen XL(®) consistently underestimated values as transducer IOP increased; however, IOP values measured with the TonoPen XL(®) were in close agreement and were less variable than those determined with the TonoVet(®) when a contact lens was applied to the cornea. Bland-Altman analysis was used to determine the lower and upper limits of agreement (TonoVet(®): -29.7 and +21.1 mmHg, TonoPen XL(®): -3.9 and +3.6 mmHg) between the two devices.
CONCLUSIONS: This study suggests that the TonoPen XL(®) is a useful tonometer for dogs wearing therapeutic contact lenses, and importantly, contact lenses would not need to be removed prior to IOP measurement.