Influence of surgical staples on radiofrequency ablation using multitined expandable electrodes.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sakuhara Yusuke, Shimizu Tadashi, Abo Daisuke, Hasegawa Yu, Kato Fumi, Kodama Yoshihisa, Shirato Hiroki
Аннотация
PURPOSE: During radiofrequency ablation (RFA), there is a risk that the multitined expandable electrode will come into contact with one of the surgical staples used to treat local recurrence after surgical operations. Our objective was to evaluate whether a surgical staple would influence the RFA of egg white using a multitined expandable electrode.
METHODS: Multitined expandable electrodes, LeVeen needles (expandable diameter 3.0 cm), were sunk into an egg white bath with (a) no surgical staple, (b) a surgical staple touching one of the tines, or (c) a surgical staple touching two of the tines simultaneously. By connecting the LeVeen needle and copper plate at the bottom of the bath, RFA was then performed on the egg whites as a substitute for human tissue. Ten egg white baths were ablated under each of conditions (a), (b), and (c), for a total of 30 sets of coagulated egg white.
RESULTS: There was no significant difference in the time from the power-on to the roll-off (i.e., the completion and shutting off of the electric circuit) or in the maximum diameter of the thermal lesion between conditions (a) and (b) or (a) and (c). However, the minimum diameter of the thermal lesion was significantly smaller in (c) compared with (a) (p < 0.01).
CONCLUSIONS: Surgical staples have the capacity to interfere with the electromagnetic field and decrease the minimum diameter of the thermal lesion in the event that a staple touches two of the tines of a multitined expandable electrode during RFA. Although the difference might be small enough to be neglected under many clinical circumstances, we recommend that, if possible, the tines not be expanded near metallic material.
METHODS: Multitined expandable electrodes, LeVeen needles (expandable diameter 3.0 cm), were sunk into an egg white bath with (a) no surgical staple, (b) a surgical staple touching one of the tines, or (c) a surgical staple touching two of the tines simultaneously. By connecting the LeVeen needle and copper plate at the bottom of the bath, RFA was then performed on the egg whites as a substitute for human tissue. Ten egg white baths were ablated under each of conditions (a), (b), and (c), for a total of 30 sets of coagulated egg white.
RESULTS: There was no significant difference in the time from the power-on to the roll-off (i.e., the completion and shutting off of the electric circuit) or in the maximum diameter of the thermal lesion between conditions (a) and (b) or (a) and (c). However, the minimum diameter of the thermal lesion was significantly smaller in (c) compared with (a) (p < 0.01).
CONCLUSIONS: Surgical staples have the capacity to interfere with the electromagnetic field and decrease the minimum diameter of the thermal lesion in the event that a staple touches two of the tines of a multitined expandable electrode during RFA. Although the difference might be small enough to be neglected under many clinical circumstances, we recommend that, if possible, the tines not be expanded near metallic material.