Effect of inter-electrode distance on bipolar intramural radiofrequency ablation.

Kovoor Pramesh, Daly Michael, Pouliopoulos Jim, Dewsnap M Barbara, Eipper Vicki, Ross David L

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

We aimed at evaluating bipolar radiofrequency ablation by correlating inter-electrode distance (ILD) with lesion dimensions and continuity.

BACKGROUND

Previous reports indicated that bipolar radiofrequency (RF) current applied to two adjacent sites in vitro, synergistically increased lesion sizes greater than that observed for unipolar RF current delivery using the same electrodes.

METHODS

Ablations were performed intramurally to ensure that each electrode surface (radius = 0.4 mm, area = 3.52 mm(2)) provided consistent contact with the myocardium. Ninety-six ablations were performed in four greyhounds using bipolar ablation needles with ILDs of 1, 2, 3, and 4 mm. An epicardial approach was used to ensure accurate positioning of the needles within the myocardium. Lesions were created using temperature-controlled RF delivery for a duration of 60 seconds to achieve 90 degrees C at the electrode proximal to the needle base. Lesion dimensions were determined histologically.

RESULTS

Increasing the ILD, decreased lesion width (P = 0.003) but increased lesion depth (P = 0.001). Lesions remained continuous with ILDs of 1-3 mm but became discontinuous at 4 mm. Energy requirements during ablation increased with increasing ILDs.

CONCLUSION

Using the above parameters (electrode radius, RF power delivery, time) during bipolar ablation, lesion continuity was critically dependent on the ILD. The maximum ILD threshold to create contiguous overlapping lesions was 3 mm. Lesions of greater width were created using shorter ILDs. Clinically, greater control over lesion dimensions can be obtained by manipulating the ILD distance.

Для цитирования: Kovoor Pramesh, Daly Michael, Pouliopoulos Jim, Dewsnap M Barbara, Eipper Vicki, Ross David L. Effect of inter-electrode distance on bipolar intramural radiofrequency ablation.. Pacing and clinical electrophysiology : PACE. 2005. DOI: 10.1111/j.1540-8159.2005.09446.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.