Vas Guide: a novel device for looping blood vessels in robot-assisted surgery.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sasaki Yutaro, Fukuta Kyotaro, Yano Tetsuhiro, Nishiyama Mitsuki, Kobayashi Saki, Minato Ryoei, Daizumoto Kei, Tomida Ryotaro, Kusuhara Yoshito, Fukawa Tomoya, Yamaguchi Kunihisa, Yamamoto Yasuyo, Furukawa Junya

Аннотация

BACKGROUND: Looping blood vessels in robot-assisted surgery remains challenging because of limitations in instrument articulation, joint length, and the lack of tactile feedback. To address this critical issue, we developed the Vas Guide, a novel device, for safe blood vessel handling. This study aims to establish the efficacy, safety, and durability of the Vas Guide.

METHODS: The Vas Guide is a flat, curved stainless-steel strip with blunt, rounded ends and a hole near the distal end. After sufficient dissection of the posterior surface of the target blood vessel, the Vas Guide with a vessel loop tied to the distal end was passed around the blood vessel in a single step using a suturing-like motion, thereby enabling secure vessel management. It is reusable and has a unit price of approximately $600. We analyzed 83 patients undergoing robot-assisted partial nephrectomy or nephroureterectomy at Tokushima University Hospital: 39 in the Vas Guide group and 44 in the conventional group. Outcomes including looping time, looping-related complication rates, such as collisions between instruments and bleeding, and additional instrument usage were recorded. The National Aeronautics and Space Administration task load index (NASA-TLX) scores were assessed.

RESULTS: The patients' characteristics were comparable between the two groups. The Vas Guide group demonstrated significantly better outcomes, including faster looping times (31 vs. 90 s, p < 0.001), fewer looping-related complications (0% vs. 30%, p < 0.001), no additional instrument usage (0% vs. 50%, p < 0.001), and better NASA-TLX scores (26.5 vs. 75.7, p < 0.001). The Vas Guide showed no usability issues, with only minor surface scratches and a slight radius change (8.60 mm to 8.69 mm).

CONCLUSION: This reusable device combines high efficacy, safety, and durability and can be implemented at a relatively low cost.