Elucidating trends in revision reverse total shoulder arthroplasty procedures: a retrieval study evaluating clinical, radiographic, and functional outcomes data.

Wiater Brett P, Baker Erin A, Salisbury Meagan R, Koueiter Denise M, Baker Kevin C, Nolan Betsy M, Wiater J Michael

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to explore relationships between damage modes in explanted reverse total shoulder arthroplasty (RTSA) components, patient and radiographic risk factors, and functional data to elucidate trends in RTSA failure.

METHODS

Fifty RTSA systems, retrieved from 44 patients, with 50 polyethylene (PE) liners, 30 glenospheres, 21 glenoid baseplates, 13 modular humeral metaphases, and 17 humeral stems, were examined for damage modes, including abrasion, burnishing, dishing, embedding, scratching, and pitting. PE liners were also analyzed for delamination and edge deformation. Charts were reviewed for patient, surgical, and functional data. Pre-revision radiographs were analyzed for scapular notching and loosening.

RESULTS

Average term of implantation was 20 months (range, 0-81 months). Metallic components exhibited abrasion, burnishing, dishing, pitting, and scratching. PE liners displayed all damage modes. Damage was exhibited on 93% of glenospheres and 100% of PE liners. Of 29 aseptic shoulders, 13 showed evidence of scapular notching and 5 of humeral loosening. There was a moderate correlation between radiographically observed implant failure or dissociation and PE embedding (r = 0.496; P < .001). There were weak and moderate correlations between scapular notching severity and PE dishing (r = 0.496; P = .006), embedding (r = 0.468; P = .010), and delamination (r = 0.384; P = .040).

CONCLUSIONS

To date, this is the largest series of retrieved RTSA components and the first to relate damage modes to radiographic and clinical data. Most damage was observed on the PE liners, on both the articular surface and rim, and glenosphere components. Correlation of retrieval findings with radiographic and clinical data may help establish predictors of prostheses at risk for failure.

Для цитирования: Wiater Brett P, Baker Erin A, Salisbury Meagan R, Koueiter Denise M, Baker Kevin C, Nolan Betsy M, Wiater J Michael. Elucidating trends in revision reverse total shoulder arthroplasty procedures: a retrieval study evaluating clinical, radiographic, and functional outcomes data.. Journal of shoulder and elbow surgery. 2015. DOI: 10.1016/j.jse.2015.06.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.