DOI 10.1016/j.arth.2009.03.010

Медицинская библиотека · The Journal of arthroplasty · 2009

Analysis of wear in retrieved mobile and fixed bearing knee inserts.

Engh Gerard A, Zimmerman Rebecca L, Parks Nancy L, Engh C Anderson

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод в очереди

Аннотация

One purported advantage for mobile bearing (MB) knee implants is reduced polyethylene wear. Twenty-three retrieved Low Contact Stress (LCS) MB and 31 Anatomic Modular Knee (AMK) fixed bearing tibial polyethylene inserts, both Enduron (Depuy, Warsaw, IN) and sterilized by gas plasma, were graded for top and backside wear using the Hood grading system. For articular surface wear, there was no statistical difference between pitting and scratching, but burnishing was twice as much for MB inserts (P = .003). For backside scores, there was a minimal amount of pitting for both, but the scratching score was twice (P = .000) and the burnishing score was 3 times greater for MB inserts (P = .000). At more than 5 years in situ, the linear wear measurements were similar for the 2 groups (mean penetration was 0.329 mm for MB and 0.320 mm for fixed bearing).

Для цитирования: Engh Gerard A, Zimmerman Rebecca L, Parks Nancy L, Engh C Anderson. Analysis of wear in retrieved mobile and fixed bearing knee inserts.. The Journal of arthroplasty. 2009. DOI: 10.1016/j.arth.2009.03.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.