[Structural damage of epidural catheters during combined spinal epidural anaesthesia: a lightmicroscopical study].

Wahlen B M, David M, Stanek A

Аннотация на английскомСтатьяПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to investigate whether there is a risk of epidural catheter damage during the advancement of the spinal needle through an epidural needle in clinical use.

METHODS

A total of 100 catheters (50 from CSE kits with a pencil-point type spinal needle and 50 from CSE kits with a Quincke type spinal needle) which had been used for routine CSE blocks were microscopically examined for any defects within the first 150 mm of the catheter. Additionally 10 unused new catheters were investigated.

RESULTS

Among 10 unused catheters 5 slight scratches were found, 92 out of 100 used catheters did not show any signs of use or scratches, 7 showed some signs of use and longitudinal scratches whereas another 1 showed a moderate scratch of less than 25% of the wall thickness. There was no difference in the prevalence of scratches between the CSE kits with pencil-point type spinal needles compared to those with Quincke-type spinal needles.

CONCLUSION

The CSE technique with either pencil-point type or Quincke-type spinal needles for subarachnoidal punctures was safe and showed no relevant epidural catheter damage.

Для цитирования: Wahlen B M, David M, Stanek A. [Structural damage of epidural catheters during combined spinal epidural anaesthesia: a lightmicroscopical study].. Der Anaesthesist. 2006. DOI: 10.1007/s00101-006-0984-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.