Медицинская библиотека · Bulletin (Hospital for Joint Diseases (New York, N.Y.)) · 1997

Severe metallosis caused by shaving down of the stem neck due to outer cup impingement of a bipolar endoprosthesis. A case report. off.

Onishi E, Iida H, Kawanabe K, Ikeda N, Nakamura T

Клинический случайСтатьяПеревод в очереди

Аннотация

The authors report on a patient with severe metallosis caused by shaving down of the stem neck due to outer cup impingement. A primary arthroplasty was performed using an Omnifit bipolar endoprosthesis (Stryker) with an acetabular bone graft fixed by ceramic screws. At the time of revision surgery the synovium was stained black. Two ceramic screws were broken, but no scratch was observed on the metal surfaces of the outer cup and the inner head. The ultra-high molecular weight polyethylene (UHMWPE) of the inner-bearing collar was worn and flat. During the operation, a groove was observed at the stem neck where the outer cup impinges. An analysis of the black pseudocapsule by scanning electron microscopy with electron prove microanalysis (SEMEPMA) revealed that the pseudocapsule contained titanium metal particles. To our knowledge, metallosis caused by the same mechanism has only been reported in four joints, which were replaced by a Zimmer BHP bipolar endoprosthesis and Howmedica bicentric cup.

Для цитирования: Onishi E, Iida H, Kawanabe K, Ikeda N, Nakamura T. Severe metallosis caused by shaving down of the stem neck due to outer cup impingement of a bipolar endoprosthesis. A case report. off.. Bulletin (Hospital for Joint Diseases (New York, N.Y.)). 1997.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.