MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mat.0000000000002264

Медицинская библиотека · ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2024

Routine Versus Selective Distal Perfusion Catheter Use in Venoarterial Extracorporeal Membrane Oxygenation.

Buda Kevin G, Robinson Emilie C, Titus Jessica, Eckman Peter M, Chavez Ivan, Cravero Ellen, Stanberry Larissa, Hryniewicz Katarzyna

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

Although current studies support the use of prophylactic distal perfusion catheters (DPCs) to decrease limb ischemia in patients on venoarterial extracorporeal membrane oxygenation (VA ECMO), methods for monitoring limb ischemia differ between studies. We evaluated the safety of a selective rather than prophylactic DPC strategy at a single center with a well-established protocol for limb ischemia monitoring. Distal perfusion catheters were placed selectively if there was evidence of hypoperfusion at any point until decannulation. All patients were followed daily by vascular surgery with continuous regional saturation monitoring. Of 188 patients supported with VA ECMO, there were no significant differences in baseline characteristics between patients with upfront, delayed, and no DPC. Thirty day mortality was highest in patients with an upfront DPC (56% in the upfront DPC group, 19% in the delayed DPC group, and 22% in the no-DPC group, p < 0.001). The incidence of major bleeding, fasciotomy, and amputation in the entire cohort was 3.7%, 3.7%, and 0%, respectively. With strict adherence to a protocol for limb ischemia monitoring, a selective rather than prophylactic DPC strategy is safe and may obviate the risks of an additional arterial catheter.

Для цитирования: Buda Kevin G, Robinson Emilie C, Titus Jessica, Eckman Peter M, Chavez Ivan, Cravero Ellen, Stanberry Larissa, Hryniewicz Katarzyna. Routine Versus Selective Distal Perfusion Catheter Use in Venoarterial Extracorporeal Membrane Oxygenation.. ASAIO journal (American Society for Artificial Internal Organs : 1992). 2024. DOI: 10.1097/mat.0000000000002264.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.