Avoiding inappropriate therapy of single-lead implantable cardioverter-defibrillator by using atrial-sensing electrodes.
Аннотация
INTRODUCTION
The single-chamber implantable cardioverter-defibrillator (ICD) can be associated with more frequent inappropriate therapies compared with dual-chamber ICDs, when they are accompanied by a simpler implantation procedure. The aim of this study was to investigate whether the use of a single-lead ICD system with atrial-sensing electrodes results in a reduction of inappropriate ICD therapy.
METHODS AND
RESULTS
The study population consisted of 212 consecutive patients, who underwent primary prophylactic single-lead ICD implantation at our institute. A ventricular lead with atrial-sensing electrodes was implanted in 77 patients (36%; Group-VDD) and a ventricular lead without atrial-sensing electrodes was implanted in 135 patients (64%; Group-VVI). Procedural and follow-up data were collected in a prospective registry. A higher prevalence of atrial fibrillation was present in Group-VDD. There were no other significant differences in patient baseline characteristics (age, sex, and other comorbidities) or follow-up period between the two groups. The operative parameters including fluoroscopic burden showed no significant differences between Group-VDD and Group-VVI. During a mean follow-up period of 697 ± 392 days, 26 patients (12%) experienced appropriate ICD therapies and 13 patients (6%) suffered inappropriate ICD therapies. The incidence of inappropriate ICD therapies in Group-VDD was significantly lower as compared to that of Group-VVI (1/77 [1%] vs 12/135 [9%]; log-rank, P = 0.028). The incidence of appropriate ICD therapies and the occurrence of device-related complications showed no significant difference between the two groups.
CONCLUSION
Single-lead ICD with atrial-sensing electrodes shows a lower incidence of inappropriate ICD therapy compared with the absence of atrial-sensing electrodes, without additional operative burden or increased complications.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.