Prospective, randomized, pathologist-blinded study of disposable alligator-jaw biopsy forceps for gastric mucosal biopsy.
Аннотация
BACKGROUND
Endoscopic biopsy forceps differ in the size and shape of the biopsy cup and the presence or absence of a needle.
METHODS
We compared four different "large cup" forceps (three with needles) designed for 2.8mm biopsy channels. A gastric antral and corpus biopsy were obtained with each. Parameters examined included: weight (mg), length (mm), orientation (poor, good), intactness (1, 2, or 3 pieces), depth (superficial, above muscularis mucosae, included muscularis mucosae), crush artefact (yes, no), and overall adequacy (inadequate, suboptimal, adequate).
RESULTS
Twenty-four patients were enrolled (191 biopsies). The median length was approximately 5mm (range 1.1-8.2mm). Histologically inadequate specimens were present in 4% with the forceps without needle compared to 16% of those with needles (P=0.061) and there were significantly fewer specimens in three or more pieces than did the forceps with needles 2.1% vs. 12.6% (P<0.05).
CONCLUSIONS
Current alligator style forceps provide a high proportion of acceptable specimens with only minor differences between brands. Forceps from one source were least preferred by endoscopy assistants and had the highest rates of inadequate biopsies and biopsies with crush artefact. Forceps without needles provide histologically acceptable samples slightly more frequently than those with needles.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.