DOI 10.1111/j.1743-6109.2008.00849.x
Endoscopic retrieval of a penile prosthesis' rear-tip extender.
Аннотация
INTRODUCTION
The renewal of penile prosthesis is a foreseeable intervention because of the limited survival time of the different implants. It usually becomes necessary simply for material fatigue after long-term regular use; in some cases, premature intervention might be necessary because of infection or other perioperative complications.
AIM: In the case presented here, the removal of a Dynaflex penile prosthesis showed a missing rear-tip extender on the right-hand tube.
MAIN OUTCOME MEASURES AND
METHODS
A cavernoscopy of the right cavernous body was performed by using a conventional urethrocystoscope. The found rear-tip extender was extracted by using alligator forceps via the cystoscopic device.
RESULTS
A new AMS CX hydraulic cavernous body prosthesis was successfully implanted. It has proven fully functional up to the current follow-up control.
CONCLUSIONS
This case shows that cavernoscopy is a gentle intraoperative procedure for diagnosing and removing material in penile prosthesis surgery.