MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tbcas.2026.3732468

Медицинская библиотека · IEEE transactions on biomedical circuits and systems · 2026

A High-Pixel-Density, Wide-DR Subretinal Prosthesis SoC with Power-Aware Light Adaptation and Charge-Recycling Local Dynamic Supply.

Eom Kyeongho, Lee Hyun-Su, Kim Minju, Im Maesoon, Lee Hyung-Min

СтатьяПеревод в очереди

Аннотация

This paper presents a 400-pixel light-adaptive retinal prosthesis (LARP) system-on-chip (SoC) that aims to support visual perception restoration. The proposed LARP SoC incorporates the power-aware light adaptation (PALA) engine and exposure-controlled light sensor (ECLS) to extend the effective light-sensing dynamic range to 55.48dB within a constrained pixel size. By combining charge-recycling auto-level-sorting multi-supply (CAM) generation with local dynamic supply (LDS) selection, the proposed system optimizes stimulation energy, with simulations showing up to a 45% reduction in stimulation energy consumption compared with conventional fixed-supply stimulation systems. The proposed pixel-clustering architecture integrates compact 80-μm × 80-μm pixels and employs a local return-electrode scheme, with the aim of potentially enhancing visual resolution by confining stimulation currents locally. Ex vivo experiments using wild-type and rd10 mouse retinas demonstrate the functionality of the LARP SoC, showing that it successfully evokes neural responses comparable to those produced by a commercial stimulator.

Для цитирования: Eom Kyeongho, Lee Hyun-Su, Kim Minju, Im Maesoon, Lee Hyung-Min. A High-Pixel-Density, Wide-DR Subretinal Prosthesis SoC with Power-Aware Light Adaptation and Charge-Recycling Local Dynamic Supply.. IEEE transactions on biomedical circuits and systems. 2026. DOI: 10.1109/tbcas.2026.3732468.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.