Endoscopic gastrostomy replacement tubes: long-term randomized trial with five silicone commercial models.
Аннотация
TRIAL
DESIGN
No analysis of the long-term performance of percutaneous endoscopic gastrostomy (PEG) replacement tubes was identified. A randomized partially blinded trial was designed hypothesizing that clinically relevant limitations of the tubes would be identified.
METHODS
Patients (N = 100, age 58.3 ± 20.7 years, 42.0% males, time with PEG 27.0 ± 22.5 months) were randomized in five parallel intervention groups, each with a tested device (Silmag(®), Bard(®), Freka(®), Kangaroo(®) and Wilson Cook(®)). Eligibility criteria included age 18-90 years, males and females, on home enteral nutrition, and the setting was a large academic hospital. Patients were allocated according to a random numbers list, and independent professionals were responsible for data collection. Primary outcome was tube longevity, calculated by Kaplan-Meier curves and Cox regression analysis. A sample of 18 was calculated based on a 10% effect size and 80% power.
RESULTS
Twenty patients were randomized in each group and all were analyzed. There was no morbidity and mortality, however tube dysfunction was common with all models. Fastener sliding occurred in 57.0% and balloon rupture in 32.0%, along with other mishaps. Best tube longevity corresponded to the Wilson Cook model (158.0 ± 10.0 days, P < 0.05), mostly on account of diminished balloon ruptures.
CONCLUSIONS
PEG replacement was safe however relatively frequent and affecting longevity, therefore longer-lasting tubes are needed. Specific attention should be given to cap, feeding line, external clamp, tube fastener; tube length, and notably balloon performance, which may be disappointing. Trial NCT01698827, CNPq Investigator Grant 302915/2011-7.