30-Day and 1-Year Outcomes of Navitor Transcatheter Aortic Valve in Low- or Intermediate-Risk Patients.

Worthley Stephen G, Giordano Arturo, Corcione Nicola, Nombela-Franco Luis, De Marco Federico, Walton Antony, Bedogni Francesco, Möllmann Helge, De Backer Ole, Leroux Lionel, Manoharan Ganesh, Tchétché Didier, Taramasso Maurizio, Li Lihua, Kuo Hai-Chien, Van Mieghem Nicolas M, VANTAGE Study Investigators

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The Navitor transcatheter aortic valve is a self-expanding, intra-annular valve indicated for patients with severe aortic stenosis (AS) at high or extreme surgical risk.

OBJECTIVES

The aim of this study was to assess the safety and effectiveness of the Navitor valve in severe AS patients at low or intermediate surgical risk.

METHODS

VANTAGE (Evaluation of TAVR Using the Navitor Valve in a Global Investigation) is a prospective, single-arm, multicenter study. The primary effectiveness endpoint was moderate or greater paravalvular leak (PVL) at 30 days; the primary safety endpoint was all-cause mortality or fatal stroke or stroke with disability at 12 months. Both endpoints were assessed against a performance goal (PG) when the sample size requirements were met. Clinical events and imaging assessments were evaluated by an independent committee and a core laboratory, respectively.

RESULTS

A total of 434 patients (203 at low risk, 231 at intermediate risk) underwent Navitor implantation between July 2021 and November 2024 across 36 sites in Europe, Australia, and Israel. The mean Society of Thoracic Surgeons Predicted Risk of Mortality scores were 1.5% and 2.6% for the low- and intermediate-risk groups. Technical success was 97.0%, with no procedural mortality. At 30 days, no patients had moderate or greater PVL (0%), which was significantly lower than the PG of 6.6% (P < 0.0001). In the first 262 patients with 12-month follow-up completed, the rate for all-cause mortality or fatal stroke or stroke with disability was 2.3%, also significantly lower than the PG of 11.3% (P < 0.0001). Sustained hemodynamic performance (mean gradient 8.0 mm Hg, effective orifice area 1.8 cm2) was seen through 12 months.

CONCLUSIONS

The Navitor valve demonstrated favorable safety and performance outcomes at 12 months, supporting its expansion to low- and intermediate-risk populations.

Для цитирования: Worthley Stephen G, Giordano Arturo, Corcione Nicola, Nombela-Franco Luis, De Marco Federico, Walton Antony, Bedogni Francesco, Möllmann Helge, De Backer Ole, Leroux Lionel, Manoharan Ganesh, Tchétché Didier, Taramasso Maurizio, Li Lihua, Kuo Hai-Chien, Van Mieghem Nicolas M, VANTAGE Study Investigators. 30-Day and 1-Year Outcomes of Navitor Transcatheter Aortic Valve in Low- or Intermediate-Risk Patients.. JACC. Cardiovascular interventions. 2025. DOI: 10.1016/j.jcin.2025.08.021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.