DOI 10.1016/j.hrthm.2024.09.016
Incidence and outcomes of cardiovascular implantable electronic device infections in patients with end-stage kidney disease.
Аннотация
BACKGROUND
End-stage kidney disease (ESKD) patients are prone to bloodstream infections that may result in a higher risk of cardiac implantable electronic device (CIED) infections.
OBJECTIVE
The objective of this study was to assess the incidence, risk predictors, management strategies, and long-term outcomes of CIED infections in ESKD patients undergoing de novo CIED implantation.
METHODS
This is a retrospective study using the United States Renal Data System. ESKD patients with de novo CIED implantation between January 1, 2006, and September 30, 2014, were included. Patients were observed until death, kidney transplantation, end of Medicare coverage, or September 30, 2015, to assess incidence of CIED infection. Management approach was determined from procedure codes for lead extraction within 60 days of CIED infection diagnosis. Patients with CIED infection were observed until December 31, 2019, to assess long-term outcomes.
RESULTS
Of 15,515 ESKD patients undergoing de novo CIED implantation, incidence of CIED infection was 4.8% during a median follow-up of 1.3 years. The presence of a defibrillator (adjusted hazard ratio [aHR], 1.48), higher body mass index (aHR, 1.01), and younger age (aHR, 0.96) were independent risk factors for CIED infection. Lead extraction occurred in only 50.71% of patients by 60 days. After propensity score matching, the 3-year mortality was higher in those who did not undergo lead extraction compared with those who did (80.3% vs 72.3%) and time to mortality was shorter (0.3 vs 0.6 year). Only 13.8% of patients underwent reimplantation with a new CIED after lead extraction.
CONCLUSION
CIED infections occur frequently in ESKD patients and are associated with a high mortality. Early lead extraction is not performed routinely but is associated with improved survival.