MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2024.01.292

Медицинская библиотека · JACC. Cardiovascular interventions · 2024

Reinterventions After CoreValve/Evolut Transcatheter or Surgical Aortic Valve Replacement for Treatment of Severe Aortic Stenosis.

Grubb Kendra J, Lisko John C, O'Hair Daniel, Merhi William, Forrest John K, Mahoney Paul, Van Mieghem Nicolas M, Windecker Stephan, Yakubov Steven J, Williams Mathew R, Chetcuti Stanley J, Deeb G Michael, Kleiman Neal S, Althouse Andrew D, Reardon Michael J

СтатьяПоддержка исследования: вне правительства СШАСравнительное исследованиеМетаанализПеревод в очереди

Аннотация

BACKGROUND

Data on valve reintervention after transcatheter aortic valve replacement (TAVR) or surgical aortic valve replacement (SAVR) are limited.

OBJECTIVES

The authors compared the 5-year incidence of valve reintervention after self-expanding CoreValve/Evolut TAVR vs SAVR.

METHODS

Pooled data from CoreValve and Evolut R/PRO (Medtronic) randomized trials and single-arm studies encompassed 5,925 TAVR (4,478 CoreValve and 1,447 Evolut R/PRO) and 1,832 SAVR patients. Reinterventions were categorized by indication, timing, and treatment. The cumulative incidence of reintervention was compared between TAVR vs SAVR, Evolut vs CoreValve, and Evolut vs SAVR.

RESULTS

There were 99 reinterventions (80 TAVR and 19 SAVR). The cumulative incidence of reintervention through 5 years was higher with TAVR vs SAVR (2.2% vs 1.5%; P = 0.017), with differences observed early (≤1 year; adjusted subdistribution HR: 3.50; 95% CI: 1.53-8.02) but not from >1 to 5 years (adjusted subdistribution HR: 1.05; 95% CI: 0.48-2.28). The most common reason for reintervention was paravalvular regurgitation after TAVR and endocarditis after SAVR. Evolut had a significantly lower incidence of reintervention than CoreValve (0.9% vs 1.6%; P = 0.006) at 5 years with differences observed early (adjusted subdistribution HR: 0.30; 95% CI: 0.12-0.73) but not from >1 to 5 years (adjusted subdistribution HR: 0.61; 95% CI: 0.21-1.74). The 5-year incidence of reintervention was similar for Evolut vs SAVR (0.9% vs 1.5%; P = 0.41).

CONCLUSIONS

A low incidence of reintervention was observed for CoreValve/Evolut R/PRO and SAVR through 5 years. Reintervention occurred most often at ≤1 year for TAVR and >1 year for SAVR. Most early reinterventions were with the first-generation CoreValve and managed percutaneously. Reinterventions were more common following CoreValve TAVR compared with Evolut TAVR or SAVR.

Для цитирования: Grubb Kendra J, Lisko John C, O'Hair Daniel, Merhi William, Forrest John K, Mahoney Paul, Van Mieghem Nicolas M, Windecker Stephan, Yakubov Steven J, Williams Mathew R, Chetcuti Stanley J, Deeb G Michael, Kleiman Neal S, Althouse Andrew D, Reardon Michael J. Reinterventions After CoreValve/Evolut Transcatheter or Surgical Aortic Valve Replacement for Treatment of Severe Aortic Stenosis.. JACC. Cardiovascular interventions. 2024. DOI: 10.1016/j.jcin.2024.01.292.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.