Brazilian journal of anesthesiology (Elsevier) · 2023

Decreasing inconsistent alarms notifications: a pragmatic clinical trial in a post-anesthesia care unit.

СтатьяPragmatic Clinical TrialНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Silveira Saullo Queiroz, Nersessian Rafael Sousa Fava, Abib Arthur de Campos Vieira, Santos Leonardo Barbosa, Bellicieri Fernando Nardy, Botelho Karen Kato, Lima Helidea de Oliveira, Queiroz Renata Mazzoni de, Anjos Gabriel Silva Dos, Fernandes Hermann Dos Santos, Mizubuti Glenio B, Vieira Joaquim Edson, da Silva Leopoldo Muniz

Аннотация

BACKGROUND: Alarms alert healthcare professionals of deviations from normal/physiologic status. However, alarm fatigue may occur when their high pitch and diversity overwhelm clinicians, possibly leading to alarms being disabled, paused, and/or ignored. We aimed to determine whether a staff educational program on customizing alarm settings of bedside monitors may decrease inconsistent alarms in the Post-Anesthesia Care Unit (PACU).

METHODS: This is a prospective, analytic, quantitative, pragmatic, open-label, single-arm study. The outcome was evaluated on PACU admission before (P1) and after (P2) the implementation of the educational program. The heart rate, blood pressure, and oxygen saturation alarms were selected for clinical consistency.

RESULTS: A total of 260 patients were included and 344 clinical alarms collected, with 270 (78.4%) before (P1), and 74 (21.6%) after (P2) the intervention. Among the 270 alarms in P1, 45.2% were inconsistent (i.e., false alarms), compared to 9.4% of the 74 in P2. Patients with consistent alarms occurred in 30% in the P1 and 27% in the P2 (p = 0.08). Patients with inconsistent alarms occurred in 25.4% in the P1 and in 3.8% in the P2. Ignored consistent alarms were reduced from 21.5% to 2.6% (p = 0.004) in the P2 group. The educational program was a protective factor for the inconsistent clinical alarm (OR = 0.11 [95% CI 0.04-0.3]; p < 0.001) after adjustments for age, gender, and ASA physical status.

CONCLUSION: Customizing alarm settings on PACU admission proved to be a protective factor against inconsistent alarm notifications of multiparametric monitors.