BACKGROUND

Transvenous lead extraction (TLE) is an important part of comprehensive lead management. The selection of tools available has expanded in recent years but data on their efficacy is limited.

OBJECTIVE

To evaluate outcomes using the TightRail™ mechanical rotating mechanical dilator sheath in comparison to excimer laser sheaths and describe factors predictive of successful extraction.

METHODS

Patients undergoing TLE at a single tertiary center (2013-2019) were included in a prospective registry. Leads targeted for extraction with either an SLS II/Glidelight™ or TightRail™ sheath were included. Outcomes were analyzed on a per-lead basis. Generalized estimating equation (GEE) models were used to assess differences in lead extraction success by extraction tool used while adjusting for nonindependence of multiple leads extracted from the same patient. Covariates included patient comorbidities, lead characteristics, and sheath size.

RESULTS

A total of 575 leads extracted from 372 patients were included. Overall success rate was 97%. TightRail™ was the first tool used in 180 (31.3%) leads with success rate of 61.7%; laser sheaths were the first tool in 395 leads (68.7%) with success rate of 67.8%. Predictors of successful extraction included lead age, lead type, and sheath sizing. Extraction success did not differ based on whether a laser or TightRail™ sheath was used (adjusted odds ratio = 0.94; 95% confidence interval = 0.59-1.50).

CONCLUSION

The TightRail™ sheath is an effective tool for TLE. Lead age, lead type, and sheath sizing were predictive of successful extraction but sheath type was not. These findings are hypothesis generating and warrant further investigation in a prospective, randomized study.

Для цитирования: Misra Satish, Swayampakala Kamala, Coons Patricia, Cerbie Claire, Guifarro Angello, Lesiczka Magdalena, Holshouser John W, Madjarov Jeko, Love Charles, Mehta Rohit. Outcomes of transvenous lead extraction using the TightRail™ mechanical rotating dilator sheath and excimer laser sheath.. Journal of cardiovascular electrophysiology. 2021. DOI: 10.1111/jce.15105.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.