BACKGROUND

Several studies compared everolimus-eluting bioresorbable scaffolds (EE-BRS) with everolimus-eluting stents (EES), but only few assessed these devices in patients with diabetes mellitus.

AIM: To evaluate the safety and efficacy outcomes of all-comer patients with diabetes mellitus up to 2 years after treatment with EE-BRS or EES.

METHODS

We performed a post hoc pooled analysis of patient-level data in diabetic patients who were treated with EE-BRS or EES in 3 prospective clinical trials: The ABSORB DM Benelux Study (NTR5447), TWENTE (NTR1256/NCT01066650) and DUTCH PEERS (NTR2413/NCT01331707). Primary endpoint of the analysis was target lesion failure (TLF): a composite of cardiac death, target vessel myocardial infarction or clinically driven target lesion revascularization. Secondary endpoints included major adverse cardiac events (MACE): a composite of all-cause death, any myocardial infarction or clinically driven target vessel revascularization, as well as definite or probable device thrombosis (ST).

RESULTS

A total of 499 diabetic patients were assessed, of whom 150 received EE-BRS and 249 received EES. Total available follow-up was 222.6 patient years (PY) in the EE-BRS and 464.9 PY in the EES group. The adverse events rates were similar in both treatment groups for TLF (7.2 vs. 5.2 events per 100 PY, p = 0.39; adjusted hazard ratio (HR) = 1.48 (95% confidence interval (CI): 0.77-2.87), p = 0.24), MACE (9.1 vs. 8.3 per 100 PY, p = 0.83; adjusted HR = 1.23 (95% CI: 0.70-2.17), p = 0.47), and ST (0.9 vs. 0.6 per 100 PY, p > 0.99).

CONCLUSION

In this patient-level pooled analysis of patients with diabetes mellitus from 3 clinical trials, EE-BRS showed clinical outcomes that were quite similar to EES.

Для цитирования: Hommels T M, Hermanides R S, Berta B, Fabris E, De Luca G, Ploumen E H, von Birgelen C, Kedhi E. Everolimus-eluting bioresorbable scaffolds and metallic stents in diabetic patients: a patient-level pooled analysis of the prospective ABSORB DM Benelux Study, TWENTE and DUTCH PEERS.. Cardiovascular diabetology. 2020. DOI: 10.1186/s12933-020-01116-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.