BACKGROUND

The photon sensitivity and spatial resolution of single-photon emission-computed tomography (SPECT) has been significantly improved by solid-state camera systems using cadmium zinc telluride (CZT) detectors. While the diagnostic accuracy of these systems is well established, there is little evidence directly comparing the prognostic utility to conventional NaI cameras.

METHODS AND

RESULTS

Retrospective analysis of patients undergoing SPECT between 2008 and 2012. Visual SPECT assessment was performed utilizing the 17-segment model to determine summed stress scores (SSS). We identified 12,830 consecutive patients, mean age 63.2 ± 13.7 and 56.1% male, 5072 of whom underwent CZT and 7758 NaI imaging. During a median follow-up duration of 7.0 years (IQR 5.5-8.2), a total of 2788 (21.7%) patients died. Compared to SSS 0, minimal perfusion abnormality (SSS 1-3) was associated with increased all-cause mortality with CZT camera (adjusted HR 1.32, P = .017) and NaI camera (adjusted HR 1.29, P = .001, interaction P = .803). Increasing stress abnormality was associated with a similar increase in risk with CZT or NaI imaging (interaction P > .500). In a propensity matched analysis, patients with normal perfusion stress perfusion assessed with a CZT was associated with decreased mortality compared to normal perfusion assessed by a NaI camera system (hazard ratio .88, 95% CI .78-.99, P = .040).

CONCLUSIONS

Increasing stress perfusion abnormality was associated with similar increase in all-cause mortality with CZT or NaI cameras. CZT and NaI camera systems provide similar risk stratification, however, normal myocardial perfusion may be associated with a more benign prognosis when assessed with a CZT camera system.

Для цитирования: Miller Robert J H, Han Donghee, Rozanski Alan, Gransar Heidi, Friedman John D, Hayes Sean, Thomson Louise, Tamarappoo Balaji, Slomka Piotr J, Berman Daniel S. CZT camera systems may provide better risk stratification for low-risk patients.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2020. DOI: 10.1007/s12350-020-02128-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.