DOI 10.1016/j.hrthm.2017.09.043
Long-term morbidity and mortality after implantable cardioverter-defibrillator implantation with procedural complication: A report from the National Cardiovascular Data Registry.
Аннотация
BACKGROUND
Long-term outcomes and predictors of mortality after implantable cardioverter-defibrillator (ICD) implantation related complication are unclear.
OBJECTIVE
The purpose of this study was to determine the risk of mortality and hospitalization after complication during ICD implantation and identify predictors of adverse outcomes.
METHODS
We performed a retrospective registry study of Medicare beneficiaries who were first-time ICD recipients enrolled in the National Cardiovascular Data Registry ICD Registry between January 2006 and March 2010. Mortality and hospitalization rates were examined using Kaplan-Meier survival analysis and multivariable Cox proportional hazards regression analysis. Covariates associated with mortality 3 years after ICD implantation complication were investigated.
RESULTS
The study cohort comprised 136,143 Medicare beneficiaries. Complications during the index hospitalization occurred in 7046 patients (5.18%), and complications within 90 days of device implantation occurred in 10,005 patients (7.34%). In adjusted analyses, complications within 90 days of implantation were associated with an increased risk of all-cause mortality at 1 year (hazard ratio [HR] 1.13; 95% confidence interval [CI] 1.05-1.20; P = .006) and 3 years (HR 1.09; 95% CI 1.05-1.13; P <.0001). These results were consistent for a combined endpoint of all-cause mortality or hospitalization. Patient, device, and hospital characteristics associated with 3-year mortality were similar between those with and those without complications.
CONCLUSION
Among Medicare beneficiaries, the occurrence of complications within 90 days of ICD implantation was associated with increased risk of all-cause mortality and all-cause mortality or hospitalization at 1 and 3 years. Patient, procedure, and hospital characteristics associated with mortality at 3 years after implantation were similar regardless of whether acute procedural complication occurred.