DOI 10.1111/pace.12049

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2012

A perioperative management algorithm for cardiac rhythm management devices: the PACED-OP protocol.

Mahlow William J, Craft Robert M, Misulia Nicholas L, Cox James W, Hirsh Jeffrey B, Snider Carolyn C, Nabers Jerrin O, Dickson Zachary A, Muenchen Robert A, Wortham Dale C

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Limited data are available regarding the perioperative management of cardiac rhythm management devices (CRMDs) exposed to intraoperative electromagnetic interference. We postulated that implementation of a simple, standardized approach to CRMD management using our own institution's Pacing And Cardioverting Electronic Devices peri-Operative Protocol (the PACED-OP protocol) would be associated with a reduction in the amount of device reprogramming without an increase in CRMD-related complications.

METHODS

Records of patients with CRMDs undergoing 497 consecutive surgical procedures were analyzed retrospectively. Roughly half (51%, n = 254) of these procedures occurred before implementation of the PACED-OP protocol, when patients were generally treated according to the American Society of Anesthesiologists' 2005 guidelines. These cases were compared to the remaining surgeries that occurred after implementation of the PACED-OP protocol. Records were screened for evidence of intraoperative CRMD malfunction that was directly associated with the use of electrocautery. Postoperative complications that could be indirectly or possibly linked to electrocautery-mediated CRMD malfunction were also identified.

RESULTS

Implementation of the PACED-OP protocol was associated with a significant reduction in the odds of device reprogramming (adjusted odds ratio [aOR] 0.19, P < 0.001). There was no direct evidence of CRMD malfunction in either cohort. The rate of postoperative complications that could be indirectly or possibly linked with electrocautery-mediated CRMD damage did not differ significantly between cohorts (aOR = 1.37, 95% confidence interval 0.56-3.3, P = 0.49).

CONCLUSION

The PACED-OP protocol implementation was associated with a significant reduction in the odds of device reprogramming without a significant difference in the odds of CRMD-related complications.

Для цитирования: Mahlow William J, Craft Robert M, Misulia Nicholas L, Cox James W, Hirsh Jeffrey B, Snider Carolyn C, Nabers Jerrin O, Dickson Zachary A, Muenchen Robert A, Wortham Dale C. A perioperative management algorithm for cardiac rhythm management devices: the PACED-OP protocol.. Pacing and clinical electrophysiology : PACE. 2012. DOI: 10.1111/pace.12049.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.