Capodanno Davide, La Manna Alessio, Di Salvo Maria Elena, Sanfilippo Alessandra, Corcos Thierry, Tamburino Corrado
Аннотация
BACKGROUND: The potential for the CATANIA (CAT) stent to be an alternative to both bare-metal stents (BMS) and drug-eluting stents (DES) has been recently demonstrated in the Assessment of The LAtest Non-Thrombogenic Angioplasty stent (ATLANTA) first-in-human study. The aim of the present study was to compare short-term outcomes of patients treated with the CAT stent with those treated with BMS.
METHODS: Based on an internal registry, the 30-day and 6-month risk-adjusted outcomes for patients who received the CAT stent (n=254) were compared against outcomes of a historical cohort of patients who received BMS (n=552) between January 2001 and December 2001.
RESULTS: At 30 days, use of BMS vs. the CAT stent resulted in borderline significant differences with respect to major adverse cardiac and cerebrovascular events (MACCE) and cardiac death or myocardial infarction. At 6 months, BMS showed a statistically significant higher adjusted risk of MACCE (HR 2.79, 95% CIs 1.20-6.48, P=.017) and no differences with respect to the subcomponent end points. The cumulative incidence of definite stent thrombosis (Academic Research Consortium defined) at 6 months was 0.39% for the CAT stent and 2.35% for the BMS.
CONCLUSIONS: This study confirms the favorable early and mid-term safety profile and the high-level efficacy of the CAT stent in the treatment of de novo coronary lesions seen in the ATLANTA trial. The use of stents with a nanothin Polyzene-F surface treatment provided improved results with respect to BMS and lower risk of acute and subacute stent thrombosis.
METHODS: Based on an internal registry, the 30-day and 6-month risk-adjusted outcomes for patients who received the CAT stent (n=254) were compared against outcomes of a historical cohort of patients who received BMS (n=552) between January 2001 and December 2001.
RESULTS: At 30 days, use of BMS vs. the CAT stent resulted in borderline significant differences with respect to major adverse cardiac and cerebrovascular events (MACCE) and cardiac death or myocardial infarction. At 6 months, BMS showed a statistically significant higher adjusted risk of MACCE (HR 2.79, 95% CIs 1.20-6.48, P=.017) and no differences with respect to the subcomponent end points. The cumulative incidence of definite stent thrombosis (Academic Research Consortium defined) at 6 months was 0.39% for the CAT stent and 2.35% for the BMS.
CONCLUSIONS: This study confirms the favorable early and mid-term safety profile and the high-level efficacy of the CAT stent in the treatment of de novo coronary lesions seen in the ATLANTA trial. The use of stents with a nanothin Polyzene-F surface treatment provided improved results with respect to BMS and lower risk of acute and subacute stent thrombosis.
Предметные рубрики MeSH
AgedAngioplasty, Balloon, CoronaryCerebrovascular DisordersCoated Materials, BiocompatibleCohort StudiesCoronary AngiographyCoronary Artery DiseaseFemaleHumansKaplan-Meier EstimateMaleMetalsMiddle AgedMyocardial InfarctionProportional Hazards ModelsProsthesis DesignRegistriesRisk AssessmentRisk FactorsStentsThrombosisTime FactorsTreatment Outcome