MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11239-005-0341-x

Медицинская библиотека · Journal of thrombosis and thrombolysis · 2004

The use of the point of care Helena ICHOR/Plateletworks and the Accumetrics Ultegra RPFA for assessment of platelet function with GPIIB-IIIa antagonists.

White Melanie M, Krishnan Rajini, Kueter Teddi J, Jacoski Mary V, Jennings Lisa K

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To evaluate a newly modified rapid platelet function analysis system (ICHOR/ Plateletworks) and to compare the results obtained with those of traditional light transmission aggregometry (LTA), and the Ultegra/RPFA system.

BACKGROUND

Anti-platelet therapy is standard of care for patients as an adjunct to percutaneous coronary intervention (PCI) or for medical management of non-ST elevation acute coronary syndromes (NSTE ACS). Recent clinical trial results suggest that the three currently approved platelet GPIIb-IIIa receptor antagonists, eptifibatide, tirofiban and abciximab, may vary in extent of inhibition of platelet aggregation (IPA) at the approved doses. Thus, pharmacodynamic evaluations of these agents to determine the extent of platelet function inhibition, especially during the periprocedural time of a cardiac intervention, are necessary. A rapid measurement method as a surrogate for LTA, the current gold standard, would be ideal in order to have the option for dose monitoring or adjustment prior to or during an intervention. The Helena ICHOR/ Plateletworks may be useful for point of care testing.

METHODS

Blood samples collected in D-Phe-Pro-Arg-chloromethyl ketone dihydrochloride (PPACK) anticoagulant were treated with increasing concentrations of eptifibatide, tirofiban or abciximab. LTA was carried out in conjunction with the ICHOR/Plateletworks, using a modified method, and Accumetrics Ultegra with RPFA cartridges.

RESULTS

This study demonstrated that platelet inhibition measured by the ICHOR/Plateletworks mirrored the level of IPA obtained with LTA. In contrast, the Ultegra system had less correlation when compared to LTA at inhibition levels < 90%.

CONCLUSIONS

Based on these data, the ICHOR/ Plateletworks utilized under modified guidelines may serve as a surrogate for LTA when rapid measurements are necessary.A rapid platelet function measurement method as a surrogate for light transmission aggregometry (LTA), the current gold standard, is ideal in order to have the option for GPIIb-IIIa antagonist dose monitoring or adjustment prior to or during a coronary intervention. A newly modified rapid platelet function analysis system (ICHOR/Plateletworks was evaluated and compared to the results obtained with traditional light transmission aggregometry (LTA), and the Ultegra/RPFA system. Blood samples collected in D-Phe-Pro-Arg-chloromethyl ketone dihydrochloride (PPACK) anticoagulant were treated with increasing concentrations of eptifibatide, tirofiban or abciximab. LTA was carried out in conjunction with the ICHOR/Plateletworks, using a modified method, and Accumetrics Ultegra with RPFA cartridges. Based on these data, the ICHOR/Plateletworks utilized under modified guidelines may serve as a surrogate for LTA when rapid measurements are necessary.

Для цитирования: White Melanie M, Krishnan Rajini, Kueter Teddi J, Jacoski Mary V, Jennings Lisa K. The use of the point of care Helena ICHOR/Plateletworks and the Accumetrics Ultegra RPFA for assessment of platelet function with GPIIB-IIIa antagonists.. Journal of thrombosis and thrombolysis. 2004. DOI: 10.1007/s11239-005-0341-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.