Real-world glycaemic outcomes observed with the use of Medtronic 780G, Tandem Control-IQ and Omnipod 5 automated insulin delivery systems.

Khan Muhammad S, Sivanantham Sathana, Farouk Lavanta, Lakshmareddy Deeksha Reddy, De Scolasticis Sara, Cresta Elisabeth, Godsland Ian, Avari Parizad, Oliver Nick, Leelarathna Lalantha, Reddy Monika

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

No randomised controlled trials have directly compared commercially available hybrid automated insulin delivery (AID) systems, and real-world comparative data remain limited. This study evaluated glycaemic outcomes across three hybrid AID systems in adults with type 1 diabetes (T1D).

METHODS

This was a retrospective, observational, single-centre study and included adults with T1D who transitioned from multiple daily injections (MDI) or non-automated insulin pump therapy to hybrid AID. Data were collected from routinely used clinical data-sharing platforms and electronic health records. Outcomes compared across systems included time in range (TIR; 3.9-10.0 mmol/L), time below range (TBR; <3.9 and <3.0 mmol/L), time above range (TAR; >10.0 and >13.9 mmol/L), glucose management indicator (GMI, %) and coefficient of variation (CV, %).

RESULTS

A total of 213 participants were included (Medtronic 780G n = 38; Tandem Control-IQ n = 81; Omnipod 5 n = 94). After adjustment for baseline TIR, diabetes duration, insulin modality and AID use duration 780G users achieved a higher TIR increase (21.1% [95% CI 18.4-23.7]) compared to Control-IQ (10.1% [3.2-17.3], p = 0.010) and Omnipod 5 (15.2% [12.9-17.5], p = 0.002), with corresponding reductions in TAR.

CONCLUSION

All three hybrid AID systems were associated with improvements in glycaemic outcomes in real-world use, supporting the role of AID systems in the management of T1D. Medtronic 780G use was associated with higher TIR increase compared with the other systems; however, these findings are based on measurements from different continuous glucose monitors between AID groups and cannot be used to infer superiority in glycaemic attainment.

Для цитирования: Khan Muhammad S, Sivanantham Sathana, Farouk Lavanta, Lakshmareddy Deeksha Reddy, De Scolasticis Sara, Cresta Elisabeth, Godsland Ian, Avari Parizad, Oliver Nick, Leelarathna Lalantha, Reddy Monika. Real-world glycaemic outcomes observed with the use of Medtronic 780G, Tandem Control-IQ and Omnipod 5 automated insulin delivery systems.. Diabetic medicine : a journal of the British Diabetic Association. 2026. DOI: 10.1111/dme.70251.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.