Audiology & neuro-otology · 2024

Comparison between the International Outcome Inventory for Hearing Aids Questionnaire and Real-Ear Measurement.

СтатьяСравнительное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Nam Gi-Sung, Lee Young Jae, Song Hansoo, Oh Jeonghyun, Cho Sung Il

Аннотация

INTRODUCTION: The international outcome inventory for hearing aids (IOI-HA) is a questionnaire widely used to assess the subjective benefits of hearing aids. This study aimed to evaluate the relationship between IOI-HA outcomes and target mismatch in real-ear measurement (REM).

METHODS: Thirty-four ears of 25 patients who had worn hearing aids were evaluated using the Korean version of the IOI-HA and REM after fitting for 2 months. Real-ear insertion gain (REIG) was measured at three different levels of input intensity - 50, 65, and 80 dB sound pressure level (SPL) - and a frequency range of 0.25-6 kHz. Factors 1 and 2 and total IOI-HA scores were compared with the mismatches of REIGs and target gains of REM.

RESULTS: Factor 1, factor 2, and total IOI-HA scores were 14.6 ± 3.5, 11.4 ± 2.2, and 25.9 ± 5.1, respectively. The averages of the difference of REIGs and target gains in REM at 50, 65, and 80 dB SPL input levels were -3.1 ± 6.7, -2.3 ± 7.2, and -3.0 ± 8.2, respectively. Factors 1 and 2 scores of the IOI-HA showed significant correlations with target mismatch in REM at 1 kHz and 0.75 kHz frequencies, respectively. Total IOI-HA scores had significant correlations with target mismatches in REM at 0.75 and 1 kHz frequencies.

CONCLUSION: IOI-HA scores correlated with target mismatch in REM at mid frequencies. The IOI-HA can be a useful screening measure for evaluating the necessity of further adjustments in hearing aids through REM at mid frequencies.