Franke-Trieger Annett, Mattheus Willy, Seebacher Josef, Zahnert Thomas, Neudert Marcus
Аннотация
OBJECTIVE: To determine the relation between stapedius reflex thresholds in cochlear implant users evoked once through direct electric stimulation on single channels (ESRT) and once through acoustic stimulation in free sound field. For comparison, stapedius reflex thresholds were obtained in free sound field in a normal-hearing control group.
DESIGN: For each participant a new ESRT-based fitting was created. Stapedius reflex thresholds were obtained for this new fitting in free sound field for different loudness adjustments. Acoustic stimuli for eliciting the stapedius reflex were narrow band noise signals covering the audiometric frequency range.
STUDY SAMPLE: N = 29 experienced CI users (34 ears) and N = 10 normal hearing listeners.
RESULTS: ESRT-based fitting resulted in different stapedius reflex behaviour compared to normal-hearing listeners. A frequency dependence was observed. Stapedius reflex thresholds decreased with increasing centre frequencies of acoustic narrow band noise stimuli. A linear relation between upper stimulation levels on the implant channels and corresponding stapedius reflex thresholds evoked in free sound field was found.
CONCLUSION: The found correlation may be a guideline for adjusting the electrical dynamic range during cochlear implant fitting. This allows the implant system to mimic the natural reflex behaviour in the best possible way and potentially avoid overstimulation.
DESIGN: For each participant a new ESRT-based fitting was created. Stapedius reflex thresholds were obtained for this new fitting in free sound field for different loudness adjustments. Acoustic stimuli for eliciting the stapedius reflex were narrow band noise signals covering the audiometric frequency range.
STUDY SAMPLE: N = 29 experienced CI users (34 ears) and N = 10 normal hearing listeners.
RESULTS: ESRT-based fitting resulted in different stapedius reflex behaviour compared to normal-hearing listeners. A frequency dependence was observed. Stapedius reflex thresholds decreased with increasing centre frequencies of acoustic narrow band noise stimuli. A linear relation between upper stimulation levels on the implant channels and corresponding stapedius reflex thresholds evoked in free sound field was found.
CONCLUSION: The found correlation may be a guideline for adjusting the electrical dynamic range during cochlear implant fitting. This allows the implant system to mimic the natural reflex behaviour in the best possible way and potentially avoid overstimulation.