Comparison of linear gain and wide dynamic range compression hearing aid circuits II: aided loudness measures.

Jenstad L M, Pumford J, Seewald R C, Cornelisse L E

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The goal of this study was to test the theoretical advantages of a single-channel wide dynamic range compression (WDRC) circuit fitted using the DSL method for increased dynamic range and normalized loudness growth.

DESIGN

Ten adolescents and young adults with moderate to severe sensorineural hearing loss were fitted monaurally with the Siemens Viva 2 Pro behind-the-ear instrument set to DSL 4.0 targets for both linear gain and WDRC processing. Threshold, upper limit of comfort and loudness growth were measured in the unaided, linear gain and WDRC conditions for warble tones, environmental sounds and speech. Twelve adult listeners with normal hearing also were tested monaurally in the unaided condition to provide normative data for comparison purposes.

RESULTS

The WDRC hearing aid provided a greater input dynamic range than the linear circuit for all stimuli. The dynamic range was normalized for more subjects with the WDRC than the linear hearing aid. In addition, exponential loudness growth functions fitted to the loudness growth data showed that, on average, loudness growth was more normalized with the WDRC hearing aid fitted to DSL[i/o] targets than the linear hearing aid fitted to DSL[i/o] targets.

CONCLUSIONS

WDRC processing, fitted using the DSL[i/o] method, has potential applications in hearing aid fittings for listeners with moderate to severe hearing loss because it provides an audible, comfortable and tolerable amplified signal across a wider range of inputs than linear gain processing, without the need for volume control adjustments.

Для цитирования: Jenstad L M, Pumford J, Seewald R C, Cornelisse L E. Comparison of linear gain and wide dynamic range compression hearing aid circuits II: aided loudness measures.. Ear and hearing. 2000. DOI: 10.1097/00003446-200002000-00007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.