International urology and nephrology · 2019

Reversed connection of cuffed, tunneled, dual-lumen catheters with increased blood flow rate maintains the adequacy of delivered dialysis despite the higher access recirculation.

СтатьяРандомизированное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kousoula Varvara, Georgianos Panagiotis I, Mavromatidis Konstantinos, Syrganis Christos, Thodis Elias, Panagoutsos Stylianos, Passadakis Ploumis

Аннотация

PURPOSE: Reversing the connection mode of permanent dual-lumen dialysis catheters results in higher access recirculation that may compromise the dialysis adequacy. The purpose of this study is to investigate the effect of reversed-connected dialysis with a higher versus a standard blood flow rate (Qb) on adequacy parameters and access recirculation.

MATERIALS AND METHODS: In a cross-over design, 46 prevalent dialysis patients with a properly functioning cuffed, tunneled, dual-lumen catheter were evaluated in three consecutive mid-week dialysis sessions. At baseline, participants were evaluated under standardized conditions (correct connection, Qb = 300 ml/min). In Phase A, dialysis was performed with reversed connection and Qb = 300 ml/min. In Phase B, dialysis was performed with reversed connection and Qb = 400 ml/min. The sequence of evaluations (Phase A and B or vise verse) was randomized. All other dialysis-related parameters were unchanged in all three occasions.

RESULTS: As expected, compared with baseline, reversed-connected dialysis in Phase A resulted in lower URR and spKt/V, and in a higher recirculation rate. Compared with baseline, reversed-connected dialysis with a higher Qb in Phase B resulted in an even higher recirculation rate, but the parameters of dialysis adequacy were not different. Increase in Qb from 300 to 400 ml/min resulted in an improvement of the dialysis adequacy (URR: 64.1 ± 7.8% vs. 70.6 ± 8.2%, P < 0.001; spKt/V: 1.22 ± 0.3 vs. 1.45 ± 0.3, P < 0.001) despite the higher recirculation rate.

CONCLUSION: This study suggests that reversed-connected dialysis with increasing Qb maintains the adequacy of the delivered dialysis despite the compensatory increase in recirculation.