DOI 10.1007/s00405-025-09475-w
Short-term audiological outcomes of the mCLIP ARC ball joint prosthesis- a retrospective multicenter study.
Аннотация
PURPOSE: This European multicentric, retrospective study aimed to analyze the short-term safety and effectiveness of the mCLIP ARC Ball Joint Prosthesis.
METHODS
Type III Tympanoplasty with implantation of a mCLIP-ARC Partial Prosthesis was performed. During patients' follow-up, ear microscopic evaluation and pure-tone audiometry were used for analysis. The post-operative pure-tone average of the frequencies 0.5, 1, 2 and 3 kHz (PTA4) were used for calculation of the air bone gap (ABG). Adverse events (AE) were evaluated during follow-up.
RESULTS
62 patients underwent implantation of the mCLIP-ARC Partial Prosthesis. 55 complete data sets were available for audiological examination. The mean follow-up time was 58 days. In two cases adverse events were recorded. The post-operative bone conduction (BC) PTA4 thresholds were stable in 96.4% of cases. The mean postoperative PTA-4 air-bone gap (ABG) was 16.5 ± 9.2 dB.
CONCLUSION
Implantation of the mCLIP ARC partial prosthesis proves to be safe and reliable. Clinical data show satisfactory audiological results and demonstrate the effectiveness of the new middle ear prosthesis with a balanced ball joint.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.