Minimizing exposure to radiation in invasive cardiology using modern dose-reduction technology: Evaluation of the real-life effects.

Faroux Laurent, Blanpain Thierry, Nazeyrollas Pierre, Tassan-Mangina Sophie, Herce Benoît, Tourneux Christophe, Metz Damien

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

We aimed to measure the reduction in the estimated dose of radiation received by patients that can be achieved using dose-reduction technology (ClarityIQ, Philips Healthcare, The Netherlands), among all patients undergoing invasive cardiology procedures.

BACKGROUND

Medical procedures remain the primary source of exposure to ionizing radiation in the general population.

METHODS

The study population comprised all patients (without exclusion criteria) undergoing invasive coronary procedures over a 1-year study period in a large referral centre equipped with two catheterization laboratories (cathlabs). Both cathlabs (A and B) were equipped with the Allura Xper FD10 imaging system (Philips Healthcare, The Netherlands), but only Cathlab B was equipped with ClarityIQ technology. The primary endpoint was the estimated total dose of radiation received by the patient, as assessed by Air Kerma (AK) and dose area product (DAP).

RESULTS

In total, 2095 invasive coronary procedures were analyzed. The patients who underwent procedures in Cathlab B received an average estimated dose that was 23% (AK) and 43% (DAP) lower than the dose received by patients undergoing procedures in Cathlab A (P < .0001). The reduction remained significant by multivariate analysis after adjustment for total X-ray time, body mass index, arterial approach, PCI of at least one lesion, sex, and patient age.

CONCLUSION

In our study, the ClarityIQ technology reduced the estimated radiation dose received by patients by 23-43%, according to the method of measurement.

Для цитирования: Faroux Laurent, Blanpain Thierry, Nazeyrollas Pierre, Tassan-Mangina Sophie, Herce Benoît, Tourneux Christophe, Metz Damien. Minimizing exposure to radiation in invasive cardiology using modern dose-reduction technology: Evaluation of the real-life effects.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2017. DOI: 10.1002/ccd.27245.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.