DOI 10.1002/rcs.1691

Медицинская библиотека · The international journal of medical robotics + computer assisted surgery : MRCAS · 2015

Design, testing and modelling of a novel robotic system for trans-oesophageal ultrasound.

Wang Shuangyi, Housden James, Singh Davinder, Althoefer Kaspar, Rhode Kawal

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Trans-oesophageal echocardiography (TOE) has been widely utilized for cardiac disease diagnosis and interventional procedure guidance. However, the TOE operator is required to manually manipulate the probe, often for long periods of time and sometimes in an X-ray environment where there is exposure to ionizing radiation.

METHODS

A novel robotic manipulation system for remote control of commercial TOE probes has been developed and tested. The system has four degrees of freedom (DOFs) and is characterized by a kinematic model. The accuracy of the model and the error propagation were analysed.

RESULTS

The prototype system was shown to exhibit the required function in terms of the mechanical reliability and range of motion. The forward kinematic model can accurately predict the trajectory of the probe tip movement. The average point-to-point errors were 2.60 mm and 3.55°.

CONCLUSIONS

Robotic assistance provided by the proposed system may improve the TOE operating environment. The proposed forward kinematic model can be further employed for automatic control. Copyright © 2015 John Wiley & Sons, Ltd.

Для цитирования: Wang Shuangyi, Housden James, Singh Davinder, Althoefer Kaspar, Rhode Kawal. Design, testing and modelling of a novel robotic system for trans-oesophageal ultrasound.. The international journal of medical robotics + computer assisted surgery : MRCAS. 2015. DOI: 10.1002/rcs.1691.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.