Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose.

Bekas Marcin, Pachocki Krzysztof A, Waśniewska Elżbieta, Bogucka Dagmara, Magiera Andrzej

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

X-ray examination is associated with patient exposure to ionizing radiation. Dose values depend on the type of medical procedure used, the X-ray unit technical condition and exposure conditions selected. The aim of this study was to determine the dose value received by patients during certain conventional radiography X-ray examinations and to assess the technical condition of medical equipment used for this purpose.

MATERIAL AND

METHODS

The study covered the total number of 118 conventional diagnostic X-ray units located in the Masovian Voivodeship. The methodology used to assess the conventional diagnostic X-ray unit technical condition and the measurement of the radiation dose rate received by patients are based on test procedures developed by the Department of Radiation Protection and Radiobiology of the National Institute of Public Health - National Institute of Hygiene (Warszawa, Poland) accredited for compliance with PN-EN 17025 standard by the Polish Centre for Accreditation.

RESULTS

It was found that 84.7% of X-ray units fully meet the criteria set out in the Polish legislation regarding the safe use of ionizing radiation in medicine, while 15.3% of the units do not meet some of them. The broadest dose value range was recorded for adult patients. Particularly, during lateral (LATl) lumbar spine radiography the recorded entrance surface dose (ESD) values ranged from 283.5 to 7827 µGy (mean: 2183.3 µGy).

CONCLUSIONS

It is absolutely necessary to constantly monitor the technical condition of all X-ray units, because it affects population exposure to ionizing radiation. Furthermore, it is essential to raise radiographers' awareness of the effects that ionizing radiation exposure can have on the human body.

Для цитирования: Bekas Marcin, Pachocki Krzysztof A, Waśniewska Elżbieta, Bogucka Dagmara, Magiera Andrzej. Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose.. Medycyna pracy. 2014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.